Home > Products > Effluent decontamination systems (EDS) > Laboratory effluent decontamination systems > MICRO10 – Laboratory effluent decontamination system

MICRO10 - Laboratory effluent decontamination system

Compact, automated electric thermal biowaste treatment with integrated energy recovery.

MICRO10 is a compact, electrically heated Effluent Decontamination System (EDS) designed for the thermal treatment of biologically contaminated liquid waste in research laboratories, biotechnology facilities, and pharmaceutical applications. With a treatment capacity of up to 80 gal/day (300 L/day), MICRO10 combines automated batch operation, controlled thermal treatment, and integrated energy recovery within a compact, self-contained system.

Its electric heating technology and built-in heat recovery eliminate the need for an external steam supply or chilled water, helping simplify laboratory utility integration.

Key specifications:

  • Capacity: Up to 80 gal/day (300 L/day)
  • Technology: Electric batch EDS
  • Treatment: 275°F (135°C) for 2 minutes, adjustable
  • Thermal lethality: F₀ ≥ 50
  • Collection tank: 66 gal (250 L), optional 106 gal (400 L)
  • Energy recovery: Integrated, no chilled water required
MICRO decontamination unit for laboratories

Why choose MICRO10 for laboratory biowaste treatment?

MICRO10 provides automated thermal decontamination for laboratories requiring greater treatment capacity than point-of-use systems, while maintaining a compact equipment footprint.

Compact plug-and-play design

MICRO10 combines liquid waste collection, thermal treatment, and automated operation in a compact unit designed to simplify laboratory installation and integration.

Fully electric operation

MICRO10 uses electric heating and does not require an external steam supply, making it suitable for laboratories without centralized steam utilities.

Integrated energy recovery

MICRO10 recovers thermal energy within its treatment process, reducing cooling utility requirements and eliminating the need for chilled water.

Automated treatment cycles

MICRO10 automatically controls and monitors treatment cycles to support consistent thermal performance while reducing manual intervention.

Flexible collection capacity

ACTINI offers MICRO10 with a standard 66 gal (250 L) collection tank and an optional 106 gal (400 L) configuration to accommodate different waste collection requirements.

MICRO10 technical specifications & benefits

Parameter

Specification

Benefits

Equipment MICRO10 Compact automated solution for laboratory biowaste treatment.
Application Laboratory liquid biowaste decontamination Supports biological liquid waste management in research and biotechnology facilities.
Technology Batch thermal treatment Provides controlled, repeatable treatment cycles.
Heating method Electric No external steam supply required.
Maximum daily capacity 80 gal/day (300 L/day) Handles higher daily volumes than compact point-of-use systems.
Effluent characteristics Aqueous / water-like Designed for water-based biological liquid waste.
Collection tank capacity 66 gal (250 L) Provides integrated waste collection before thermal treatment.
Optional collection tank 106 gal (400 L) Offers additional collection capacity for variable waste generation.
Treatment vessel capacity 2.6 gal (10 L) Enables controlled batch thermal treatment.
Standard treatment cycle 275°F (135°C) for 2 minutes Delivers controlled thermal exposure to support biological inactivation.
Treatment settings Adjustable Allows adaptation to specific treatment requirements.
F₀ ≥ 50 Provides a measurable thermal lethality parameter to support treatment verification.
Electrical supply 120 V / 60 Hz / single-phase Uses the specified US single-phase electrical supply.
Electrical power 3 kW / 32 A Provides electric heating without an external steam supply.
Energy recovery Integrated Recovers heat within the treatment process.
Cooling requirements No chilled water required Simplifies facility utility connections.
Standard dimensions (L × W × H) 35 × 43 × 83 in (880 × 1,100 × 2,100 mm) Compact footprint for laboratory installation.
Dimensions with 400 L tank 36 × 46 × 88 in (920 × 1,170 × 2,240 mm) Provides increased collection capacity with a defined equipment footprint.

Confirm installation requirements and available configurations during project engineering.

How does the MICRO10 effluent decontamination system work?

MICRO10 uses automated batch thermal treatment with integrated energy recovery to process biologically contaminated liquid waste.

1. Liquid waste collection

MICRO10 collects biologically contaminated liquid waste in its integrated storage tank before treatment.

2. Automated batch treatment

The system transfers the collected effluent into its thermal treatment vessel and applies controlled electric heating and holding-time conditions.

3. Treatment monitoring

MICRO10 monitors treatment parameters during each cycle to support consistent operation and verification of the thermal process.

4. Heat recovery and discharge

The system uses integrated heat recovery during the treatment process. After cooling, MICRO10 discharges the treated effluent through the designated outlet.

Treatment parameters, effluent compatibility, and discharge requirements must match the facility’s application and operating conditions.

Tecnical schema

MICRO10 configuration and available options

Programme de traitement

Standard collection tank

MICRO10 includes a 66 gal (250 L) collection tank for biological liquid waste storage before treatment.

Ajout - scalable - cuve

Extended collection capacity

ACTINI offers an optional 106 gal (400 L) collection tank to accommodate increased storage needs or variable waste generation.

ISO - environment

Integrated heat recovery

MICRO10 incorporates energy recovery within its thermal treatment process and operates without a chilled water connection.

supervision

Automated control and monitoring

The system uses automatic PLC control and process monitoring to support consistent treatment cycles and routine operation.

Typical MICRO10 applications

MICRO10 is designed for laboratories and small-scale biotechnology facilities requiring automated treatment of moderate volumes of aqueous biological liquid waste.

Potential applications include:

  • Research and development laboratories
  • Biotechnology laboratories
  • Pharmaceutical research facilities
  • Laboratories with centralized liquid waste collection points
  • Facilities upgrading from smaller point-of-use decontamination equipment

ACTINI USA evaluates effluent composition, biological treatment requirements, daily treatment capacity, and installation constraints to confirm equipment suitability.

MICRO decontamination unit for laboratories
Tecnical schema

Need a compact laboratory EDS with higher treatment capacity?

Whether you’re equipping a new laboratory or upgrading an existing biological liquid waste treatment process, ACTINI USA helps you evaluate MICRO10’s suitability for your application.

Our engineering team supports equipment selection, installation planning, commissioning, and qualification activities according to your project requirements.

Frequently asked questions about MICRO10 (FAQ)

What is the maximum treatment capacity of MICRO10?

MICRO10 treats up to 80 gal/day (300 L/day) of biological liquid waste. Actual treatment capacity depends on operating conditions and the selected thermal treatment parameters.

Does MICRO10 require steam or chilled water?

No. MICRO10 uses electric heating and integrated energy recovery. It does not require an external steam supply or chilled water connection.

What treatment temperature does MICRO10 use?

MICRO10’s standard treatment program operates at 275°F (135°C) for 2 minutes. ACTINI can adjust treatment settings according to the selected configuration and validated process requirements.

What is the difference between MICRO10 and KUBE?

Both systems use automated electric batch thermal treatment.

KUBE is designed for compact point-of-use applications and treats up to 26 gal/day (100 L/day). MICRO10 provides greater treatment capacity, reaching 80 gal/day (300 L/day), with integrated energy recovery and a larger collection tank.

Can MICRO10 handle effluents containing solids or viscous materials?

MICRO10 is designed for aqueous, water-like biological liquid waste. For effluents containing significant solids or viscous materials, ACTINI can evaluate alternative treatment technologies, including DIRECTSTEAM.